054gs-750x500
  Exceed ACR Accreditation Requirements Test the full range of standard diagnostic ultrasound probes (2 MHz to 18 MHz) Detachable water wells allow for testing curvilinear and endocavity probes Ensure over ten years of reliable use through reinspection and repair services The Model 054GS Ge ...Read more



  • Description
  • Specifications

 

Exceed ACR Accreditation Requirements

  • Test the full range of standard diagnostic ultrasound probes (2 MHz to 18 MHz)
  • Detachable water wells allow for testing curvilinear and endocavity probes
  • Ensure over ten years of reliable use through reinspection and repair services

The Model 054GS General Purpose Ultrasound Phantom contains gray scale targets and improved sensitivity targets that exceed the phantom requirements outlined in the ACR accreditation program.

The phantom is made of proprietary Zerdine® hydrogel polymer, which has been formulated to provide tissue mimicking properties including compatibility with harmonic imaging. To maximize phantom lifetime, this gel is contained in a rugged ABS plastic housing with a Saran-based laminate membrane.

 

Key Tests with Model 054GS

  • Uniformity
  • Depth of Penetration
  • Beam Profile/ Focal Zone/ Lateral Response Width
  • Vertical Distance Measurement
  • Horizontal Distance Measurement
  • Axial and Lateral Resolution
  • Elevational Resolution
  • Contrast Resolution
  • Grayscale Contrast Sensitivity
  • Dead Zone Assessment

 

 

Dimensions 17.8 cm x 12.7 cm x 20.3 cm
Weight 4.1 kg
Housing Materials ABS Plastic
Membrane Saran-based laminate
Tissue-Mimicking Material Zerdine® solid elastic hydrogel
   
Model 057GS Includes  
1x General Purpose Ultrasound Phantom
1x Detachable Protective Cover
1x Detachable Water Well (1 cm deep)
1x Detachable Endocavity Well
1x Carry Case
1x User Guide
1x Certificate of Compliance

 

Zerdine® Properties

  • Freezing point: 0° C
  • Melting point: Above 100° C
  • Speed of Sound: 1540 m/s
  • Other: Compatible with harmonic imaging

 

Vertical Distance Group

  • Number of targets: 10
  • Wire diameter: 100-micron, nylon monofilament
  • Depth range: 1 to 18 cm
  • Spacing: 20 mm

 

Horizontal Distance Group

  • Number of groups: 7
  • Wire: 100-micron, nylon monofilament
  • Depth: 9 cm
  • Spacing: 20 mm respectively

 

Near Field Group

  • Number of targets: 6
    Wire Diameter: 100 microns, nylon monofilament
    Depth range: 1 to 6 mm
    Distance b/w Targets: 1 mm 

 

Axial-Lateral Resolution Groups

Wire diameter: 80 microns

  • Group 1 Depth: 3 cm
    • Axial separation: 4, 3, 2, 1, 0.5 & 0.25 mm
    • Lateral separation:    4, 3, 2, 1, 0.5 & 0.25 mm
  • Group 2 Depths: 11 cm
    • Axial separation: 5, 4, 3, 2 & 1 mm
    • Lateral separation: 5, 4, 3, 2 & 1 mm

 

Anechoic Cylinders

  • Number of targets: 5
  • Diameter of targets:  8 mm
  • Depth of Targets: 4, 7, 10, 13, & 16 cm
  • Contrast: Anechoic, Cyst-like

 

Gray Scale Targets

  • Number of targets: 6
  • Contrast: -6 dB, -3 dB, +3 dB, +6 dB & hyperechoic  (about 12 dB) with respect to background
  • Depth: 4 cm
  • Diameter: 8 mm