- Description
- Specifications
This Image Quality Indicator is used for testing all important parameters of CR scanner systems according to ASTM E 2445-05, ISO 16371-1 and EN 14784-1 standards. It provides means to measure: spatial resolution, unsharpness, contrast, MTF, laser beam jitter, scanner slipping and shading.
Technical data:
- Dimensions: 350 x 430 x 19 mm
- Made of PMMA (Lucite)
- T shaped brass target (laser beam jitter, MFT check, blooming / flare)
- Pro-NDT DuplexWireIQI test object (basic spatial resolution, unsharpness)
- BAM snail made of lead and low absorbent material (central beam alignment)
- Two converging line pairs resolution patterns (Pro-Res MAM ConeType)
- Three Ø19 x 0.3 mm (EL, EC, ER) measuring points (shading correction)
- Cassette positioning locator
- Homogeneous 0.5 mm Al strip (scanning slipping, shading)
- Two cm / inch rulers (linearity check)
- Three contrast sensitivity gauges made of Al, Cu and stainless steel with four holes of different depth
- Calibration / test certificate according to ASTM E 2445-05, EN 14784-1 and ISO 16371-1
10-101 - version with one Pro-NDT DuplexWireIQI
Product features:
|