QA StereoChecker™

qasc_750x500
  Filmless CyberKnife® QA System Designed to be used by both physicists and therapists, this easy method for complete daily CyberKnife QA enables improved frequency and depth of QA practices, potentially improving clinical results.     MLC QA CyberKnife Daily QA The refined MLC Gard ...Read more



  • Description
  • Specifications

 

Filmless CyberKnife® QA System

Designed to be used by both physicists and therapists, this easy method for complete daily CyberKnife QA enables improved frequency and depth of QA practices, potentially improving clinical results.

 

 

MLC QA

CyberKnife Daily QA

The refined MLC Garden Fence test provides the absolute leaf position results against Accuray’s four MLC test criteria – two for overall MLC bank accuracy and two for individual leaf accuracy. MLC leaf position errors as small as 0.2mm are detected.

Iris QA

CyberKnife Daily QA

Easily acquire scan-quality data every morning for radial profiles, field width, penumbra, and flatness. This test detects field size variations as small as 0.1mm.

PANDA QA

CyberKnife Daily QA

Replace the film-based Accuray AQA test with one that is faster, easier, and more detailed. Iris and fixed cone PANDA tests measure the translational and rotational offsets in 3 dimensions in addition to the total magnitude. This test detects offsets as small as 0.1mm.

Monthly Linac QA

Easily perform single field analysis on MLC, jaw and fixed cone collimated linac fields. View inline and crossline profiles for field sizes up to 15 x 15 cm together with test metrics for field width, penumbra, flatness and symmetry.

Monthly CyberKnife QA

Expand upon Daily CyberKnife QA with single field analysis tests. Test metrics and profiles are automatically calculated, displayed, saved and trended for Iris, MLC and fixed cone collimated fields.

Manual Mode for Linac and CyberKnife

Acquire images with outstanding 200 micron resolution, independent of linac systems, for effortless DICOM export and analysis in software applications.

Build on a solid foundation

Installation and setup have been completely reimagined, providing users with a robust, secure, and intuitive experience as a locally-hosted web application.

Simplified & optimized

QA StereoChecker is designed with therapists in mind, as they are key to efficient morning QA being done. No hassle device setup and uncomplicated software provide the means to successfully integrate CyberKnife QA into the therapist’s daily workflow.

Analysis

Test results are quickly displayed upon image acquisition. They are also accessible through quick links Review and Trending workspaces.

Trends

Version 2.0 now provides full trending of all test metrics through graphical and tabular displays making it easier than ever understand machine performance issues so they may be addressed in a timely manner, securing optimal machine uptime.

Reports

QA session reports and longitudinal reports are promptly generated with the click of a button.

 

 

General Specifications

  • Dimensions: 38.7 cm L x 32.7 cm W x 4.9 cm H
  • Photons: 60Co to 15 MV

 

Sensor

  • Total Pixel Number: 1024 x 1024
  • Active Pixel Number: 1000 x 1000
  • Pitch: 200 µm
  • Total area: 204.8 x 204.8 mm2

 

Electronics

  •  ADC: 16 bit, 1 MSps

 

Detector

  • Dynamic Range: > 88 dB
  • Scintillator: Gd2O2S:Tb

 

Software Specifications

  • Operating system: Windows® 10 Professional
  • Processor: Intel Core (At least dual-core recommended)
  • Memory: 8 GB RAM (16 recommended)
  • Hard Drive: 2 GB Free (20 GB Recommended)
  • Screen resolution: 1024 x 768 or higher
  • Ports: RJ45 (Ethernet)