Contact
Industrial
Physics
Radiotherapy
Radiology
Ultrasound
Home

Pro-NM Linear Source Module

pro-nm_linear_source_module_750x500
  This module can be used as a standalone in air or in water if mounted in the Pro-NM Performance cylinder. It can be used to evaluate changes of radius-of rotation on spatial resolution, spatial resolution measurement in air and in water, quantitative evaluation of reconstruction filters and scatt ...Read more



  • Description
  • Specifications

 

This module can be used as a standalone in air or in water if mounted in the Pro-NM Performance cylinder. It can be used to evaluate changes of radius-of rotation on spatial resolution, spatial resolution measurement in air and in water, quantitative evaluation of reconstruction filters and scatter compensation methods.

 

Technical data:

  • Insert diameter: 186 mm
  • Diameter of line sources: 1 mm
  • Spacing of line sources: 75 mm
  • Useful height of line sources: 70 mm
  • Stopcocks with luer connection allow for easy and safe filling and draining of line sources

 

 

Product features:

  • Complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM  (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • The manual provides detailed guidelines for carrying out each test, results assessment and registration