- Description
- Specifications
The phantom for performance testing of mammography X-ray units. It has 49 holes that generate subtle changes in contrast. It can be used to detect small variations in system’s performance.
Technical data:
- Dimensions: 62.7 x 62.7 x 62.7 mm
- Typical contrast for mammographic energies: 6.60, 4.20, 2.60, 1.70, 1.00, 0.65, 0.41%
- Hole depth (7 columns): 0.853, 0.533, 0.332, 0.208, 0.129, 0.080, 0.050 mm
- Diameter of holes (7 rows): 4.292, 2.524, 1.485, 0.873, 0.513, 0.302, 0.177 mm
- Optimized for digital units
- Includes a rotating support plate
- Comfortable carrying case
Product features:
|